Hotline: 0904.754.399

Công ty TNHH Công nghệ ATP Việt Nam

Tiếng việtEnglish


Tropicool Incubator
View Full-Size Image


Tropicool Incubator

( Công ty TNHH Công nghệ ATP Việt Nam )



The following should be read in conjunction with information covering the entire BIOCELL Series of incubators, as details unique to the Tropicool Variants are described below.

To enable laboratory technicians to deal costeffectively with bacteriological culturing in the tropics, Contherm has developed the Tropicool system. This entirely new thermoelectric technology represents a significant advance on traditional refrigeration techniques - and gives the Contherm Biocell “Tropicool” variant incubators a new dimension in quality and performance. The Tropicool thermoelectric system is a reliable, economical and environmentally-friendly means of cooling. The heart of this system and serving as the active cooling element, is a solid state, semiconductor
device which contains no moving parts or plumbing, yet is very effective at transferring heat energy. With low power consumption, operating temperatures down to 22°C at an ambient of 30ºC, for instance, can be achieved (see graph). The Biocell “Tropicool” variant incubators enable quality incubating in the tropics, and B.O.D. work in temperate regions. Plate and sample drying problems are greatly reduced with the "Tropicool" system. Higher than "normal" humidities are maintained due to the low temperature difference between the system's cooling fins and the interior air, thus elevating the dew point. The unique fan scroll also contributes to this capability, providing a dynamically-balanced true one-pass air circulation system. An easy-to-set and read digital LED display allows the laboratory professional to operate these incubators up to +50°C with a set-point accuracy of 0.1°C without problems. A range of seven attractive and dependable models is available, to satisfy the diverse needs of laboratory technologists. The Biocell “Tropicool” variant incubators are manufactured in a quality production environment, and are built to NZS6200, AS3100 Electrical Safety Standards.
More technical information is available from the Contherm Techspec library. The words 'Contherm' and 'Biocell' are the property of Contherm Scientific Ltd. *Warranty covers components and labour for 1 year. Full warranty applies only with regard to published pricelist. The information contained in this brochure is correct at time of printing but is subject to change without notice.

MODEL

1050T

1100T

1150T

1200T

1300T

1400T

1800T

Nominal Capacity

50 litres

100 litres

150 litres

200 litres

300 litres

400 litres

800 litres

External Dimensions

 

 

 

 

 

 

 

(mm) Height

655

770

855

915

1215

1515

1550

Width

520

590

675

785

785

785

1380

Depth

630

730

800

800

800

800

800

Internal Dimensions

 

 

 

 

 

 

 

(mm) Height

335

500

580

640

940

1240

1240

Width

385

455

540

650

650

650

1220

Depth

340

440

485

505

505

505

505

Shelves Supplied

2

3

3

3

4

5

5

Load: Watts

360W

360W

510W

510W

510W

510W

900W

Net Weight

42kg

62kg

70kg

90kg

112kg

122kg

275kg

Packed Weight

46kg

66kg

74kg

98kg

124kg

135kg

325kg

Operating Temperature

See graph showing performance envelope ( Maximum +50°C )

Operating Ambient

10° to 35°C

Temporal Variation

±0.2°C

±0.3°C

Spatial Variation

±0.4°C

±1.2°C

Power Supply

230 - 240Volt 50hz 10Amp Grounded Supply

Controller

ZP21 Digital LED Readout

Timer

1 minute to 99 hours 59 minutes

Options

Option X - Castors. Option Z - Cable Port. Option E - RCD.

 







You may also be interested in this/these product(s)

Incubator
Incubator
Incubator INE
Incubator INE
Incubator INB
Incubator INB
Galaxy® R Series CO2 Incubators
Galaxy® R Series CO2 Incubators



Last Updated: Monday, 11 December 2023 05:49