Product Maintenance Guides

pH Electrode Care

1351
com_content.category
(2 votes, average 4.50 out of 5)
pH Electrode Care4.50 out of 50 based on 2 voters.


 

Electrode Calibration

Since glass pH electrodes measure H+ concentration relative to their reference half-cells, they must be calibrated periodically to ensure accurate, repeatable measurements. Our wide selection of commercial pH calibration buffers include solutions standardized against NIST-certified pH references for calibrating meters with resolution up to 0.001 pH.

Read more...

Effects of Chemicals on Labware

1350
com_content.category
(2 votes, average 5.00 out of 5)
Effects of Chemicals on Labware5.00 out of 50 based on 2 voters.


Chemicals may affect the weight, strength, color, dimensions, flexibility, and surface appearance of labware. The basic models of interaction that cause these changes are: (1) chemical attack on the polymer chain, with resultant reduction in physical properties, including oxidation; reaction of functional groups in or on the chain; and depolymerization; (2) physical change, including absorption of solvents, resulting in softening and swelling of the plastic; permeation of solvent through the plastic;

Read more...

Six Ways to Simplify Fluid Transfer

1349
com_content.category
(2 votes, average 5.00 out of 5)
Six Ways to Simplify Fluid Transfer5.00 out of 50 based on 2 voters.


 

To pump viscous fluids 
- Choose tubing at least two sizes larger than the flow size required.
- Keep motor speed <300 rpm; the slower the motor, the better the flow.
- Use adjustable occlusion pumps: 
 Over-occlude to prime pump;
 reduce occlusion for longer life.

Read more...

Part 2: Troubleshooting pH Electrodes

1348
com_content.category
(2 votes, average 5.00 out of 5)
Part 2: Troubleshooting pH Electrodes5.00 out of 50 based on 2 voters.


 

Getting the optimal performance from your pH electrodes is important. Several factors can influence an electrode’s functionality, including drifting readings, slow response times, and environmental influences. Find out how to determine what may be affecting your electrode—and what you can do about it.

DRIFTING READINGS

Drifting readings can occur if you are using a gel-filled electrode under the following conditions: when measuring a relatively large volume of sample, measuring samples at slightly colder temperatures, or testing fairly clean water. For reference, read the following article: http://www.coleparmer.com/techinfo/techinfo.asp?htmlfile=pHPureWater.htm

Read more...

How to Calibrate an RTD or Platinum Resistance Thermometer (PRT)

1347
com_content.category
(2 votes, average 5.00 out of 5)
How to Calibrate an RTD or Platinum Resistance Thermometer (PRT)5.00 out of 50 based on 2 voters.


 

Introduction

There are two types of calibrations applicable to PRTs—characterization and tolerance testing. The type of calibration to perform is determined by the way in which the UUT is to be used and the accuracy required by the user.

Characterization is the type of calibration in which the unit under test (UUT) resistance is determined at several temperature points and the data are fitted to a mathematical expression.

Read more...